Lee, S., and Feng, X., and Liu, M.H., and Zhang, G.X., (1997) A Crystal Spectrograph and its Application in Plasma Focus. Jurnal Fizik Malaysia, 18 (1). pp. 1-10. ISSN 0128-0333 Full text not available from this repository. AffiliationsNational Institute of Education. Nanyang Technological University. School of Science. AbstractA new compact crystal spectrograph (size 4x4x7 cm³) for soft x-ray measurement in a plasma focus facility was designed and studied. The spectrograph has an adjustment system with 5 degrees of freedom making it easy to align and use. The spectrograph provides spatially resolved spectnim in one dimension over 0.1-5 nm wavelength range. A TIAP crystal with different form (flat or curve) was used to obtain spatially resolved spectra of Ne and Ar ions from DPF plasma focus. Such a spectrograph is also suitable to be used coupled to soft x-ray streak camera or x-ray CCD detectors. | Item Type: | Journal |
|---|